• Главная
  • Продукты
  • Производители
  • О компании
  • Вакансии
  • Акции и предложения
  • О холдинге
  • EMC Partner серии MIG и MIG0603OMI

    Дополнительная информация


    Test System Feature
    • Slow oscillatory waves 100kHz, 1MHz
    • Pulse repetition up to 400Hz for slow waves
    • Amplitude up to 3000V
    • Output impedance 200 ohm
    • Fast oscillatory waves 3MHz, 10MHz, 30MHz
    • Pulse repetition 5kHz
    • Burst duration 1ms to 50s
    • Burst repetition 20ms to 999s
    • Amplitude up to 4000V
    • Output impedance 50 ohm
    • Automatic coupling path selection
    • Impulse parameters defined at CDN output
    • Parameter ramp feature
    • Electronic polarity change
    • Semiconductor switches
    • Compact design
    • High degree of automation without software
    • 2 year warranty
    User Benefits
    The technical excellence and many unique features of damped oscillatory generators translate directly into benefits for the user:
    • Do not obsolete existing equipment. EMCP generators already in use can be enhanced to cover the new requirements.
    • Controlled from MIG-OS-OS1 or MIG0603OSI.
    • One coupler for all AC power tests.
    • CDN3000-06-25 can be used for all damped oscillatory wave, electric fast transient, surge and ringwave testing.
    • Generator with balanced outputs.
    • Positive and negative outputs for injection in balanced data lines.
    • CDN2000-06-25 can be used for slow damped oscillatory wave, electric fast transient, surge and ringwave testing. In addition, up to 4 data lines can be tested.
    • Save operator time with the automated test routines and test report facility.
    • Unparalleled reliability and system up-time.
    Generators
    One system is available with expansion capabilities to cover all damped oscillatory test requirements.
    Many of the generators include more than one impulse type, making them ideal for laboratory applications.
    For all generators, the most significant test parameters can be programmed as fixed values on the instrument front panel, or using the ramp function to change parameters during a test. Impulse voltage level, synchronisation angle and polarity can all be programmed using this feature.

    - MIG-OS-OS1
    The “classic” damped oscillatory wave generator for IEC, EN and ANSI applications. The slow waves are available through either balanced differential or grounded outputs. The 100kHz and 1MHz waveforms can be programmed up to 3000V. This generator is ideal for applications requiring only the slow waves.

    - MIG0603OSI
    A combination of slow wave generator up to 3000V, combination wave generator up to 6000V/3000A and 1.2/50μs voltage impulse up to 6000V with 0.5J energy content. All impulses can be used for IEC, EN and ANSI applications.

    - MIG-OS-OM
    A combination of the “classic” damped oscillatory wave generator for IEC, EN and ANSI applications. The “slow waves” are available through either balanced differential or grounded outputs. The 100kHz and 1MHz waveforms can be programmed up to 3000V. Apart from the voltage impulses, this generator has outputs for magnetic field testing using the 1m x 1m antenna.

    - MIG0603OMI
    A combination of 100kHz & 1MHz voltage slow waves, 1.2/50μs voltage impulse up to 6000V with 0.5J energy content and magnetic field testing for AC, impulse and damped oscillatory waves. All can be performed using the same 1m x 1m antenna.

    - MIG-OS-MF
    Can be used to perform magnetic field tests for damped oscillatory waves (100kHz and 1MHz). Maximum output 150A/m using the 1m x 1m antenna.
     Long duration testing is made easier by use of the EMC PARTNER software packages Using either GENECS or TEMA software, the units can be programmed, automatically started and test reports generated.
     The compact design enables many different test standards to be performed using only a single unit. A broad range of  accessories enable testing to many additional applications.
     Special configurations are available to meet unique customer requirements.





    Damped Oscillatory Waves Magnetic Field
    with MF1000-1
    Oscillation frequency 100kHz
    Antenna dimension 1m x 1m
    Current range 1 up to 150A (100A/m)
    with MF1000-2 (100kHz)
    Oscillation frequency 100kHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 130A (120A/m)
    with MF1000-2 (1MHz)
    Oscillation frequency 1MHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 92A (60A/m)
    MIG0603OMI
    Damped Oscillatory Waves (DOW)
    Oscillation frequency 100kHz & 1MHz
    Voltage range 0.2 up to 3kV
    Source impedance 200 ohms
    Voltage increment 1V steps
    Damping rate 50% between 3rd and 6th periods
    Pulse rise time 75ns
    Burst repetition at 100kHz up to 50Hz
    Burst repetition at 1MHz up to 500Hz
    Polarity Positive, negative, alternating
    Damped Oscillatory Waves Magnetic Field
    with MF1000-1
    Oscillation frequency 100kHz
    Antenna dimension 1m x 1m
    Current range 1 up to 150A (100A/m)
    with MF1000-2 (100kHz)
    Oscillation frequency 100kHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 130A (120A/m)
    with MF1000-2 (1MHz)
    Oscillation frequency 1MHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 92A (60A/m)
    Impulse Magnetic Field
    with MF1000-1
    Waveshape 8/20ps
    Antenna dimension 1m x 1m
    Current range 200 up to 2200A (1500A/m)
    with MF1000-2
    Waveshape 8/20ps
    Antenna dimension 1m x 2.6m
    Current range 100 up to 1500A (1000A/m)
    AC Magnetic Field (50 & 60Hz)
    with MF1000-1
    Antenna Dimension 1m x 1m
    Current Range 0.8 up to 160A/m
    with MF1000-2
    Antenna dimension 1m x 2.6m
    Current range 0.4 up to 80A/m
    with MF1000-3
    Antenna dimension 1m x 1m
    Current range 100 up to 1100A/m
    Combination Wave 1. 2/50ps (8/20ps)
    Voltage range 0.5 up to 6kV
    Voltage increment 1V steps
    Short circuit current 0.25 up to 3000A
    Source impedance 2 ohms
    Energy at 6kV 220J
    Pulse interval 8 seconds at max. voltage
    Polarity Positive, negative, alternating
    Programmable parameter ramps Voltage, synchronisation, polarity
    Surge IEC 60255-5 0.5J 500 Ohm
    Waveform at no load No load = R > 100Ohm
    Rise time 1.2ps
    Time to half value 50ps
    Adjustable voltage range 500V - 6000V
    Polarity pos. / neg. / alternate
    Test Energy Generator voltage Impedance Impulse Voltage capacitor Waveform OC Current  Waveform  SC
    0.5kV 0.5J 500Ohm 4.0pF 0.5kV 1.2/50ps 1A 1.2/50ps
    1kV 0.5J 500Ohm 1.0pF 1kV 1.2/50ps 2A 1.2/40ps
    3kV 0.5J 500Ohm 110nF 3kV 1.2/50ps 6A 1.2/25ps
    5kV 0.5J 500Ohm 40nF 5kV 1.2/50ps 10A 1.2/10ps
    MIG-OS-MF
    Damped Oscillatory Waves Magnetic Field
    with MF1000-1
    Oscillation frequency 100kHz
    Antenna dimension 1m x 1m
    Current range 1 up to 150A (100A/m)
    with MF1000-2 (100kHz)
    Oscillation frequency 100kHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 130A (120A/m)
    with MF1000-2 (1MHz)
    Oscillation frequency 1MHz
    Antenna dimension 1m x 2.6m
    Current range 1 up to 92A (60A/m)

    External Three Phase Couplers
    Damped oscillatory wave testers can be extended with automatic or manual three-phase coupling networks. The CDN2000-06-25 and CDN3000A-06-25 can also be used for EFT, CWG surge and ring wave. Coupling path selection is either from the MIG firmware, from GENECS and TEMA software or manually on the CDN front panel (manual version only). The coupling networks fulfill the requirements laid down in the IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-12 (ring wave), IEC61000-4-18, ANSI C37.90 and ANSI C62.41 standards.
    CDN2000-06-25
    Manual three phase CDN with line voltages L to N=280V and L to L=415V, line current 25A per phase. Coupling path selection for EFT, Damped Oscillatory, Surge and Ring. Differential input for Damped Oscillatory Waves. (MIG-OS-OS1, MIG0603OSI, MIG0603OMI, MIG-OS-OM)
    Note:
    CDN2000-06-25 can be used for testing on up to four data lines using the 100kHz and 1MHz oscillatory waves.
    CDN2000-06-25 OPTION 480V
    Extends three phase CDN for line voltages L to N = 280V and L to L = 480V. Applicable for damped oscillatory wave, surge (CWG), EFT/burst and ring wave impulses.
    CDN3000A-06-25
    Three phase CDN with line voltages L to N=280V and L to L=480V, line current 25A per phase. Automatic coupling path selection for EFT, Damped Oscillatory, Surge and Ring wave. (MIG-OS-OS1, MIG0603OSI, MIG0603OMI, MIG-OS-OM, MIG-OS-OM-EXT)
    Note:
    Common Mode coupling only.
    Application:
    IEC 61000-4-18 fast waves de-/coupling
    Magnetic Field Testing
    Applicable standards are IEC 61000-4-8 (AC), IEC 61000-4-9 (IMPULSE) and IEC61000-4-10 (DOW).  MIG damped oscillatory wave testers can be extended by the MF1000 coil antennas.  MF1000 antennas are used to generate magnetic fields when connected to the MIG outputs. The following levels can be reached:
    Coil dimensions AC magnetic fields (50/60Hz) Impulse magnetic fields (8/20ps) DOW magnetic fields (100kHz & 1MHz)
    MF1000-1 1m x 1m 1 up to 130A/m 0.1 up to 1.5kA/m 1 up to 150A/m
    MF1000-2 1m x 2.6m 1 up to 110A/m 0.1 up to 1.1kA/m 1 up to 130A/m
    MF1000-3 1m x 1m 0.3 up to 1kA/m n/a n/a

    Номер в
    госреестре
    Наименование СИ Обозначение типа СИ Изготовитель Срок свидетельства
    или заводской номер
    Не подлежит сертификации

    Запросить цену можно по телефону или заполнив эту несложную форму.

    ФИО:

    Компания:

    Контактные данные: